Tiswati, Kandi Ayu (1999) Perbandingan respon dosis radiasi pada Film Badge dan dosimeter Elektronik dengan alat ukur standard Dosimeter Farmer. Undergraduate thesis, FMIPA UNDIP.
PDF Restricted to Repository staff only 1755Kb | ||
| PDF 17Kb | |
| PDF 347Kb | |
| PDF 503Kb | |
| PDF 354Kb | |
| PDF 896Kb | |
| PDF 382Kb | |
PDF Restricted to Repository staff only 462Kb | ||
| PDF 331Kb | |
| PDF 333Kb | |
| PDF 502Kb |
Abstract
Telah dilakukan penelitian untuk membandingkan penerimaan dosis radiasi antara film badge dengan dosimeter elektronik dengan alas ukur standard dosimeter Farmer. Penelitian ini dilakukan dengan memberikan paparan radiasi gamma 137Cs sebesar 10, 20, 30, 40, 50, 75, 100, dan 200 mR terhadap film badge, dosimeter elektronik, dan alat ukur standard pada jarak sate meter dan kondisi penyinaran yang sama. Data dianalisis dengan menggunakan uji ANOVA dan dilanjutkan dengan uji Duncan. Hasil penelitian menunjukkan ada perbedaan nyata untuk besar dosis radiasi pada film badge, dosimeter elektronik dan alat ukur standard. Perbedaan film badge terhadap alat ukur standard sebesar 8,98 %, dan dosimeter elektronik terhadap alat ukur standard sebesar 8,70 %. Perbedaan dosis ini cukup memadai untuk tingkat proteksi radiasi, dengan perbedaan kurang dari 20 %. A research to compare the receiving radiation dose between film badge with electronic dosemeter by farmer dosemeter as standard measurement device has been performed. The research was curried out with given radiation gamma 137Cs exposure of 10, 20, 30, 40, 50, 75, 100 and 200 Ira to film badge, electronic dosemeter and standard measurement device on same distance on one meter and exposure conditions. The data were analyzed with using of ANOVA test and continued with Duncan test. The research result shown the real differences for radiation dose on film badge, electronic dosemeter and standard measurement device. The difference film badge with standard measurement device was about 8,98 % and electronic dosemeter with standard measurement device about 8,70 %. This dose difference was enaugh for radiation protection level, with difference less than 20 %.
Item Type: | Thesis (Undergraduate) |
---|---|
Subjects: | Q Science > QC Physics |
Divisions: | Faculty of Science and Mathematics > Department of Physics |
ID Code: | 30466 |
Deposited By: | Mr UPT Perpus 1 |
Deposited On: | 31 Oct 2011 12:41 |
Last Modified: | 31 Oct 2011 12:41 |
Repository Staff Only: item control page