Harmoko , Udi (1996) Pengaturan penguat dan penganalisa salur tunggal untuk pencacahan radio formaka Co - 57. Undergraduate thesis, FMIPA UNDIP.
PDF Restricted to Repository staff only 1899Kb | ||
| PDF 16Kb | |
| PDF 345Kb | |
| PDF 437Kb | |
| PDF 396Kb | |
| PDF 671Kb | |
| PDF 677Kb | |
PDF Restricted to Repository staff only 573Kb | ||
| PDF 349Kb | |
| PDF 337Kb | |
| PDF 473Kb |
Abstract
Telah dilakukan penelitian pengaturan penguat dan penganalisa salur tunggal untuk sistem pencacah gamma pada satu pilihan sumber radiofarmaka Co-57. Sebagai sumber kalibrator_dipakai sumber Cs-137. Pulsa sumber Co-57 dikuatkan 264,2 kali dan pulses sumber Cs-137 dikuatkan 34,1 kali, sehingga puncak serap¬an fotolistrik kedua sumber tersebut terletak pada kanal yang sama. Dengan mengatur jendela penganalisa salur tunggal pada pencacahan Co-57 yaitu batas ambang bawah (LLD) = 2,8 Volt dan batas'ambang atas (ULD) = 3,2 Volt. Sedangkan untuk sumber cs-137, batas ambang bawah (LLD) = 2,5 Volt dan batas ambang atas (ULD) = 3,2 Volt, sehingga diperoleh jumlah cacah masing - masing sumber (Cs-137 dan Co-57) adalah sama untuk setiap pencacahan kedua sumber tersebut saat aktivitas keduanya sama, yang merupakan prinsip dasar penelitian ini. Dari pengujian menggunakan sumber Co-57 yang berak-tivitas 1,36 pCi didapatkan hasil yang mendekati harga sesungguhnya yaitu 1,36 + 0,02 pCi. It have been researched the gain of amplifier and single channel analyzer's window for Gamma Counter system on a Radiopharmacitical Co-57 choice and the Cs-137 as calibration source. The Co-57 source pulse have to be amplified by 264.2 -times and Cs-137 by 34.1 times to locate both peak area on the same channel. The single channel analyzer's window for counting Co-57 -be arranged by Low Level Discriminator at 2,8 Volt and Upper Level Discriminator at 3.2 Volt. Furthermore for counting Cs-137, The Low Level Discriminator at 2.5 Volt and Upper Level Discriminator at 3.2 Volt. So it would reached for principle of the instrument designing, that every counts both source (Co-57 or Cs-137) wich have the same activity will produce the same counting rate too. The result of designing have been tested by Co-57 source at 1.36 pCi produced the same limited really value was 1.36 ± 0.02 pCi.
Item Type: | Thesis (Undergraduate) |
---|---|
Subjects: | Q Science > QC Physics |
Divisions: | Faculty of Science and Mathematics > Department of Physics |
ID Code: | 30287 |
Deposited By: | Mr UPT Perpus 1 |
Deposited On: | 26 Oct 2011 07:58 |
Last Modified: | 26 Oct 2011 07:58 |
Repository Staff Only: item control page