Ketut Umiati, Ngurah Ayu and Nur, Muhammad and Darmini, Darmini (2003) Pengaruh Kecepatan Penguatan Lembar Penguat Terhadap Densitas Radiograf. BERKALA FISIKA, 6 (3). pp. 63-70. ISSN 1410 - 9662
| PDF - Published Version 148Kb |
Abstract
A research about influence of speed of the intensifying screen to density on a radiograph in diagnose has been done. The influence speed of the intensifying screen to density has been curried out by measuring the exposure values at the same densities for intensifying screen high speed, medium speed and low speed with step wedge as object. The result of this research indicated that the intensifying screen medium speed is needed 2 times greater exposure value than intensifying screen high speed, while intensifying screen low speed is needed 1,5 times greater exposure value than intensifying screen medium.
Item Type: | Article |
---|---|
Uncontrolled Keywords: | Exposure factor (V.i.t), density, radiation dose. |